Towards Estimating Normal and Shear Interface Pressures in Prosthetic Sockets via Least Squares and Mechanics Modeling
通过最小二乘法和力学建模估计假肢接受腔中的法向和剪切界面压力
Axel González Cornejo、Tianhao Yu、Chi Hwan Lee、Edgar Bolívar-Nieto · arXiv:2606.04222 · 2026-06-02 · cs.RO
假肢接受腔的装配仍然主要是手动和迭代的,客观的装配指标仍然有限。部分挑战是残肢-接受腔界面缺乏长期的现实压力数据。传统的压力传感器很容易随着时间的推移而发生漂移,并且仅捕获插座内稀疏位置的正常压力,缺少生物力学分析的关键组成部分:剪切力。尽管一些传感器可以报告法向界面应力和剪切界面应力,但由于测量串扰,这些组件通常难以解耦。一种潜在的前进道路是开发可以增强可用 m 的模型。
中文为机器翻译节选,以英文原文为准。
English abstract
Prosthetic socket fitting remains largely manual and iterative, and objective fit metrics are still limited. Part of the challenge is the lack of long-term real-life pressure data at the residual limb--socket interface. Traditional pressure sensors are prone to drift over time, and capture only normal pressures at sparse locations within the socket, missing a critical component for biomechanical analysis: shear. Although some sensors can report both normal and shear interface stresses, these components are often difficult to decouple because of measurement crosstalk. One potential path forward is to develop models that can augment available measurements. This work introduces a testbed to evaluate model performance under sparse pressure sensing using two complementary validation signals: (i) the global wrench (\ie, total forces and moments expressed in an orthonormal frame) transmitted…