Comparing Commercial Depth Sensor Accuracy for Medical Applications

比较医疗应用的商用深度传感器精度

Pit Henrich、Maximilian Weiherer、Franziska Hansen、Bernhard Egger、Franziska Mathis-Ullrich · arXiv:2606.13028 · 2026-06-11 · cs.RO

论文自动收录 · 未精读PDF ↗arXiv 原文 ↗
壹摘要

深度估计有许多医学和外科应用。我们使用触针采样参考对猪骨样本、猪腹部样本和硅胶肾模型上的四个深度传感器进行基准测试。这些对象包含几个现实世界的挑战,包括均匀表面、镜面表面和次表面散射。比较包括距离大约 50 厘米的立体、结构光和飞行时间传感器。具体来说,Intel RealSense D405(Intel RealSense,美国)、PMD Flexx2(pmdtechnologies,德国)、Stereolabs ZED 2i(Stereolabs,法国)和 Zivid 2M+ 60(Zivid,挪威)

中文为机器翻译节选,以英文原文为准。

English abstract

Depth estimation has numerous medical and surgical applications. We benchmark four depth sensors on a porcine bone specimen, a porcine belly specimen, and a silicone kidney phantom using stylus-sampled references. These objects contain several real-world challenges, including homogeneous surfaces, specular surfaces, and subsurface scattering. The comparison includes stereo, structured-light, and time-of-flight sensors at a distance of approximately 50 cm. Specifically, the Intel RealSense D405 (Intel RealSense, United States), PMD Flexx2 (pmdtechnologies, Germany), Stereolabs ZED 2i (Stereolabs, France), and Zivid 2M+ 60 (Zivid, Norway) are compared. The Zivid 2M+ 60 performed best across all objects and metrics considered in this work. The ZED ranked second for real tissue, but last on the phantom.

贰引用
Pit Henrich、Maximilian Weiherer、Franziska Hansen 等 5 人. Comparing Commercial Depth Sensor Accuracy for Medical Applications. arXiv:2606.13028 (2026-06-11). https://arxiv.org/abs/2606.13028v3