Interaction Dynamics Modeling and Predictive Control for Safe Steerable Catheter--Tissue Interaction
安全可转向导管的相互作用动力学建模和预测控制--组织相互作用
Yongyan Cao · arXiv:2607.20939 · 2026-07-23 · eess.SY
可操纵导管是包括射频消融在内的心脏电生理学 (EP) 手术的主要工具,其中尖端必须精确定位在目标组织上,同时保持受控、稳定的接触。因此,中央控制问题不仅仅是尖端跟踪和力调节;它是\emph{导管-组织相互作用动力学}的调节。交互状态必须对尖端如何相对于组织移动、持续摩擦和接触力如何偏置该运动以及安全限制如何重塑物理上允许的运动进行编码。现有方法调节这些相互作用动态
中文为机器翻译节选,以英文原文为准。
English abstract
Steerable catheters are the primary tool for cardiac electrophysiology (EP) procedures including radiofrequency ablation, where the tip must be positioned precisely at target tissue while maintaining controlled, stable contact. The central control problem is therefore not merely tip tracking and not merely force regulation; it is the regulation of \emph{catheter--tissue interaction dynamics}. The interaction state must encode how the tip moves relative to tissue, how persistent friction and contact forces bias that motion, and how safety limits reshape what motion is physically allowable. Existing methods regulate these interaction dynamics through different mechanisms. Classical impedance control~\cite{hogan1985} shapes the tip port as a virtual mechanical impedance Z(s) = M d s² + D d s + K d, providing passive compliance without an explicit contact model. Three complementary…