Capstan-driven Continuum Surgical Robot: Design, Modeling, and Perception

绞盘驱动的连续手术机器人:设计、建模和感知

Gang Zhang、Yufu Qiu、Junyan Yan、Wenhui Zeng、Wenlong Lu、Shing Shin Cheng · arXiv:2608.13396 · 2026-08-13 · cs.RO

论文自动收录 · 未精读PDF ↗arXiv 原文 ↗
壹摘要

形状和力传感长期以来一直是紧凑型绞盘驱动的连续手术机器人开发的关键瓶颈,这主要是由于难以获取受限绞盘组件内的电缆张力信息。为了克服这些挑战,本文提出了一种基于驱动-感知协同设计概念的集成设计-建模-传感方法。驱动系统的电机安装支架中引入了柔顺元件,能够在电缆反作用力作用下发生微变形,从而无需占用紧凑的绞盘空间即可实时测量电缆张力。致地址

中文为机器翻译节选,以英文原文为准。

English abstract

Shape and force sensing have long been critical bottlenecks in the development of compact capstan-driven continuum surgical robots, primarily due to the difficulty of obtaining cable tension information within the confined capstan assembly. To overcome these challenges, this paper presents an integrated design-modeling-sensing approach based on the concept of actuation-perception co-design. A compliant element is introduced into the motor mounting bracket of the drive system, enabling micro-deformation under the cable reaction force and thereby allowing real-time cable tension measurement without occupying the compact capstan space. To address the modeling complexity arising from unconventional joint configurations introduced by the spatial cable routing strategy, a parallel computation framework based on a multibody short-thick-beam model is proposed, which captures shear effects in…

贰引用
Gang Zhang、Yufu Qiu、Junyan Yan 等 6 人. Capstan-driven Continuum Surgical Robot: Design, Modeling, and Perception. arXiv:2608.13396 (2026-08-13). https://arxiv.org/abs/2608.13396v1
叁同分类新篇