Shaft-Configuration-Adaptive Catheter Tip Position Estimation via Motor-History Conditioned Residual Learning
通过运动历史条件残差学习进行轴配置自适应导管尖端位置估计
Peihan Zhang、Michael C. Yip、Ankur Kapoor、Young-Ho Kim · arXiv:2609.26304 · 2026-09-22 · cs.RO
肌腱驱动的连续机械手广泛应用于医疗应用,其中准确的尖端位置估计对于精确导航和仪器定位至关重要。然而,患者解剖结构和程序设置强加了任务相关的未知轴配置,而摩擦、松弛和顺应性引入了滞后,使得尖端估计具有挑战性。本文提出了一种运动历史条件门控循环单元(GRU)残差估计器,用于在不直接轴配置传感的情况下进行三维导管尖端估计。首先,应用初始多向扫描策略来校准几何导管
中文为机器翻译节选,以英文原文为准。
English abstract
Tendon-driven continuum manipulators are widely used in medical applications, where accurate tip-position estimation is essential for precise navigation and instrument positioning. However, patient anatomy and procedural setup impose task-dependent unknown shaft configurations, while friction, slack, and compliance introduce hysteresis, making tip estimation challenging. This paper presents a motor-history-conditioned gated recurrent unit (GRU) residual estimator for three-dimensional catheter tip estimation without direct shaft-configuration sensing. First, an initial multidirectional sweep strategy is applied to calibrate a geometric catheter model backbone, and encode the motor-angle and drive-torque response into a shaft-configuration context vector. During subsequent motion, the context conditions a GRU that predicts a task-space residual correcting this backbone, relying on motor…