Drape-Compatible Tool-Tip Localization for Hand-Held Laparoscopic Instruments via UWB Carrier-Phase Ranging and Trocar-Constrained Geometry
通过 UWB 载波相位测距和套管针约束几何形状对手持式腹腔镜器械进行悬垂兼容工具提示定位
Jinseok Lee、Dongho Yee、Minsung Kim、Younghoon Noh、Seonho Shim、Juahn Oh、Yechan Seo、Jiyul Lee 等 11 人 · arXiv:2609.39129 · 2026-09-30 · cs.RO
外科手术机器人策略需要知道每个仪器的工作端相对于相机和另一个仪器的位置,但腹腔镜手术只留下内窥镜视频,而覆盖的仪器本身隐藏了每条光路。我们仅通过距离和 IMU 来估计工具提示。这些距离是通过仪器手柄侧天线节点之间的超宽带载波相位来测量的:每个手持仪器一个,内窥镜一个,全部位于无菌屏障之外。以内窥镜天线为参考。两根仪器天线携带六个坐标,而三对天线则提供三个距离
中文为机器翻译节选,以英文原文为准。
English abstract
A surgical robot policy needs to know where each instrument's working end sits relative to the camera and to the other instrument, yet a laparoscopic operation leaves only the endoscope video, and a draped instrument hides every optical path on itself. We estimate the tool tips from distances and an IMU alone. The distances are measured by ultra-wideband carrier phase between antenna nodes on the handle side of the instruments: one per hand-held instrument, one on the endoscope, all outside the sterile barrier. Take the endoscope antenna as the reference. The two instrument antennas carry six coordinates while the three pairs give three distances, so the problem is short by three, and averaging cannot close that gap because what is missing is information, not precision. Carrier phase adds one more unknown per pair, an integer that leaves distance fixed only modulo lambda/2 = 23.1 mm.…