Autonomous Robotic Navigation for Endovascular Brain-Computer Interface Access

用于血管内脑机接口访问的自主机器人导航

Harry Robertshaw、Weijie Qi、Nikola Fischer、Alejandro Granados、Thomas C. Booth、Sam E. John · arXiv:2610.03537 · 2026-10-02 · cs.RO

论文自动收录 · 未精读PDF ↗arXiv 原文 ↗
壹摘要

血管内脑机接口(BCIs)避免了开颅手术,但需要通过解剖学上可变的脑静脉进行精确的设备输送。这项工作首次展示了用于脑静脉系统血管内 BCI 接入的体外自主机器人导航。使用一种训练解剖结构的几何增强,对 Soft Actor-Critic 控制器进行计算机模拟训练,以执行从右颈内静脉到上矢状窦的两个连续任务。在训练解剖学和解剖学上看不见的保留模型中对导航进行了评估,超过 250 个模拟实验和 5 个透视引导的体外实验

中文为机器翻译节选,以英文原文为准。

English abstract

Endovascular brain-computer interfaces (BCIs) avoid craniotomy but require precise device delivery through anatomically variable cerebral veins. This work presents the first demonstration of in vitro autonomous robotic navigation for endovascular BCI access in the cerebral venous system. Soft Actor-Critic controllers were trained in silico for two sequential tasks spanning the right internal jugular vein to the superior sagittal sinus, using geometric augmentation of one training anatomy. Navigation was evaluated in a training anatomy and an anatomically unseen hold-out model over 250 in silico episodes and five fluoroscopy-guided in vitro robotic runs per task-anatomy condition, comprising 1,000 simulated episodes and 20 physical runs overall. Task recurrent predictors were also evaluated for online identification of impending navigation failure. In silico success rates for Tasks A and…

贰引用
Harry Robertshaw、Weijie Qi、Nikola Fischer 等 6 人. Autonomous Robotic Navigation for Endovascular Brain-Computer Interface Access. arXiv:2610.03537 (2026-10-02). https://arxiv.org/abs/2610.03537v1
叁同分类新篇