壹摘要

眼科显微外科手术中的术中场景感知受到仪器遮挡的限制:当工具进入高放大倍率视野时,它会对外科医生和处理视频的机器人辅助系统的感知组件隐藏组织。生成修复可以移除遮挡物,但通过合成来替换丢失的图像内容,这在视觉来源很重要的情况下是不合适的。许多视频修复程序还依赖于未来的帧,无法在线操作。我们提出了 TransGHOST,一种因果参考引导的重建管道,其非生成安全模式实时运行。每个恢复值都是obt

中文为机器翻译节选,以英文原文为准。

English abstract

Intraoperative scene perception in ophthalmic microsurgery is limited by instrument occlusion: when a tool enters the high-magnification field, it conceals tissue from both the surgeon and the perception components of robot-assisted systems that process the video. Generative inpainting can remove the occluder but replaces the missing image content with synthesis, which is unsuitable when visual provenance matters. Many video inpainters also depend on future frames and cannot operate online. We present TransGHOST, a causal reference-guided reconstruction pipeline whose non-generative safe mode operates in real time. Every restored value is obtained by resampling a recorded frame using geometric transforms fitted only to visible regions. The system provides per-pixel confidence and provenance and marks unsupported pixels as abstentions. We evaluate TransGHOST on a new veterinary clinical…